HARRICK PLASMA

Oxygen Service Vacuum PumpsĀ 

    Our Plasma Cleaners require a vacuum pump with a minimum pump speed of 1.4 m3/hr (0.83 ft3/min) and an ultimate total pressure of 200 mTorr (0.27 mbar) or less

    Features

    Suitable for pumping concentrated oxygen gas as well as nonreactive or inert gases (e.g. Air, N2 or Ar)

    Does not use hydrocarbon oil (the combination of concentrated oxygen and hydrocarbon oil is an explosion hazard)

    PDC-OPD (115V); PDC-OPD-2 (230V)
    Dry Oxygen Service Pump
    Uses no oil or fluid
    Pumping speed (50/60 Hz): 6.2 m3/hr
    Ultimate total pressure (gas ballast closed; 50/60 Hz): 15 mTorr
    Ultimate total pressure (gas ballast open; 50/60 Hz): 38 mTorr
    Motor power (50/60 Hz): 260 W
    Weight: 58 lbs
    Dimensions: 12" H x 17" W x 11" D
    Pump inlet: includes hose, hose clamps, hose adaptor, centering ring and swing clamp to connect plasma cleaner to pump inlet
    PDC-OP-OLT

    Optional pump outlet connection to conduct pump exhaust to a vent hood
    Includes 9 ft. length exhaust hose (other lengths may be specified) 

    PDC-OPE (115V); PDC-OPE-2 (230V)
    Economy Dry Oxygen Service Pump
    Uses no oil or fluid
    Pumping speed (50/60 Hz): 3.0/3.6 m3/hr
    Ultimate total pressure: 250 mTorr (pump manufacturer specification)
      - In our lab tests, ultimate pressures of <100 mTorr were attained.
    Motor power (50/60 Hz): 120 W
    Weight: 21 lbs
    Dimensions: 7" H x 14" W x 5.5" D
    Pump inlet: includes hose, hose clamps, hose adaptor, centering ring and swing clamp to connect plasma cleaner to pump inlet
    PDC-OPE-OLT

    Optional pump outlet connection to conduct pump exhaust to a vent hood
    Includes 9 ft. length exhaust hose (other lengths may be specified) 

    PDC-OPF (115V); PDC-OPF-2 (230V)
    Fomblin-Based Oxygen Service Pump (not including Fomblin Fluid)
    Pumping speed (50/60 Hz): 3.3/3.9 m3/hr
    Ultimate total pressure (gas ballast closed): 1.5 mTorr
    Ultimate total pressure (gas ballast open): 90 mTorr
    Motor power (50/60 Hz): 250/300 W
    Weight: 48 lbs
    Dimensions: 9" H x 17" W x 6" D
    Integrated inlet check valve
    Pump inlet: includes hose, hose clamps, hose adaptor, centering ring and swing clamp to connect plasma cleaner to pump inlet
    Pump outlet: includes oil mist filter to filter oil vapor from pump exhaust
    PDC-FMB (1kg; 0.532L)
    Fomblin Fluid for Fomblin-Based Vacuum Pump PDC-OPF, PDC-OPF-2
    PDC-OP-OLT
    Optional pump outlet connection to conduct pump exhaust to a vent hood Includes 9 ft. length exhaust hose (other lengths may be specified)